Standard turnaround: 7–10 business days
Rush available
On-site or mail-in
Pressure range: up to 15,000 psi
NIST-traceable where applicable
Rush service available
PRESSURE INSTRUMENT COVERAGE / BY INSTRUMENT TYPE
Pressure instruments we service, grouped by type.
Pressure instruments range from electronic transmitters to mechanical gauges and switches, so calibration starts with knowing how the device senses and reports pressure. These are the instrument types we most often calibrate, verify, and repair.
T / 01
ELEC
Electronic
Transmitters, transducers, and sensors that convert pressure into an analog or digital signal.
PT
Pressure Transmitters
ST
Smart Transmitters
TD
Pressure Transducers
PS
Pressure Sensors
T / 02
MECH
Mechanical
Analog and vacuum gauges that read pressure directly, plus deadweight testers used as references.
PG
Pressure Gauges
VG
Vacuum Gauges
DW
Deadweight Testers
T / 03
DIG
Digital & Recording
Digital gauges, indicators, and recorders that display or log pressure readings.
DI
Digital Pressure Gauges
IN
Pressure Indicators
RC
Pressure Recorders
T / 04
DIFF
Differential
Instruments that measure the difference between two pressures, along with their valve manifolds.
DP
Differential Pressure Transmitters
DG
Differential Pressure Gauges
MF
Manifolds
T / 05
LOOP
Switches & Loops
Pressure switches, alarms, and complete loops from the sensing point to the control system.
SW
Pressure Switches
AL
Pressure Alarms
LP
Complete Pressure Loops
Not sure which type your instrument is?
Send us the make and model. We’ll identify the instrument type and recommend an appropriate calibration or verification approach for your application.



FIELD EQUIPMENT / TRANSMITTERS, GAUGES, SWITCHES & PRESSURE LOOPS
FIELD DIAGNOSTICS / COMMON SYMPTOMS
What is your pressure instrument telling you?
Many service requests start with a reading that doesn’t look right. These are common pressure instrument symptoms and the areas a technician typically checks. The actual cause is confirmed through troubleshooting before any adjustment or repair.
SYMPTOM INDEX / S 01–06
MEASURED
EXPECTED
DEVIATION
S / 01
DRIFT
Readings drift over time
CAN POINT TO
Gradual sensing element changes, temperature or overpressure effects, or zero and span settings that no longer match the application.
WHAT WE CHECK
Zero and span against a reference
Sensor and diaphragm condition
As-found readings, increasing and decreasing
S / 02
OUTPUT
Output doesn’t match the local gauge or display
CAN POINT TO
Range or scaling differences between the transmitter and the receiving system, analog output drift, or loop wiring and termination problems.
WHAT WE CHECK
Analog output against the display
Range and scaling at both ends
Loop wiring and termination
S / 03
NOISE
Erratic or unstable reading
CAN POINT TO
Pulsation from pumps or valves, trapped air or liquid in impulse lines, grounding or shielding problems, or electrical interference.
WHAT WE CHECK
Impulse lines and process connections
Grounding, wiring, and shielding
Damping and signal quality at the transmitter
S / 04
ZERO
Reads zero while pressure is present
CAN POINT TO
A closed or plugged isolation valve or impulse line, lost loop power or wiring faults, or a failed sensor or transmitter.
WHAT WE CHECK
Valves, manifolds, and impulse lines
Loop power, wiring, and connections
Sensor and transmitter function
S / 05
DRIFT
Pressure switch trips at the wrong point
CAN POINT TO
Setpoint drift, a worn or sticking mechanism, a deadband that no longer suits the process, or contact and wiring problems.
WHAT WE CHECK
Trip and reset points against a reference
Setpoint and deadband adjustment
Contacts and wiring
S / 06
LEVEL
Tank level reads high or low
CAN POINT TO
A zero or span offset, a changed mounting reference, buildup on the sensor, or specific gravity and range settings that no longer match the installation.
WHAT WE CHECK
Level reading against a reference
Sensor mounting and condition
Range, specific gravity, and scaling settings
Seeing a different symptom?
Describe what the instrument is doing and include the make and model. A technician can help determine whether troubleshooting, calibration, or repair is the right next step.
CALIBRATION CHECK
0–100 PSI RANGE
01 REFERENCE
75.0
PSI
02 TEST POINTS
%
PSI
mA
0%
0.0
4.00
25%
25.0
8.00
50%
50.0
12.00
75%
75.0
16.00
100%
100.0
20.00
03 TRANSMITTER OUTPUT
16.00
mA
4
8
12
16
20
ZERO · SPAN · LINEARITY
COMPARED TO REFERENCE
Pressure calibration for the complete measurement loop, not just the gauge.
Our pressure instrument service can include zero and span checks, reference comparison at multiple points on increasing and decreasing pressure, range and scaling, analog output verification, switch trip and reset checks, configuration, wiring, and complete-loop performance. Service is available on site or through our mail-in calibration lab for eligible equipment.
Calibration and verification
Zero, span, and output checks
On-site and in-lab options
Detailed calibration documentation
Resolve drift, signal faults, and configuration errors across pressure systems
Inspection, troubleshooting, repair, and recalibration can help restore dependable operation, or clarify when sensor, transmitter, gauge, or switch replacement is the practical option.
Pressure Instrument Service Questions
Clear answers about calibration, verification, schedules, supported instruments, and service documentation.